Which measuring range does 20 %–100 % refer to in the 2003 edition?
Applies to: ISO 6789:2003 · Updated 2026-11-17
Technical review: Alston Hsiao (Chieh-Jen Hsiao), CTO, STAND
20 %–100 % means from one fifth of the tool’s maximum torque value up to the full value; it is the specified measuring range covered by the requirements and test methods of the 2003 edition. It is not “20 % of each reading,” nor does it mean “the guarantee extends as far as the scale is drawn.”
Find the maximum value first, then convert to actual torque
The following is teaching data: suppose an adjustable wrench has a maximum torque value of 150 N·m. Under the 2003 range rule, the lower limit is 150 × 20 % = 30 N·m and the upper limit is 150 × 100 % = 150 N·m, so the specified measuring range is 30–150 N·m.
If a job requires 45 N·m, 45 ÷ 150 × 100 % = 30 %, which lies within this range. If it requires 18 N·m, 18 ÷ 150 × 100 % = 12 %, which is not within this specified range. This is only a judgment of position, not a measurement result, and it does not mean that every point within the range has been verified as conforming.
The same working torque represents a different percentage on a tool with a different maximum value. Before deciding, therefore, check this tool’s maximum value and unit; do not rely only on its shape or on habits from using other tools.
Keep measurement points and the whole range separate
The 2003 edition generally arranges measurement points at 20 %, approximately 60 %, and 100 % of the maximum value. Type II Class B and E fixed-adjustment tools, however, are measured at the nominal value or set value, so the general three-point arrangement cannot be forced onto them.
If a tool has no scale mark exactly at 20 %, the 2003 edition specifies that the deviation be determined at the scale mark immediately below it instead. This deals with a measurement point that cannot fall exactly on a scale mark; it cannot be read as extending the whole specified measuring range downward. If a report’s low point is slightly below 20 %, first check whether this provision applies before concluding that the measurement was wrong.
A visible low-end scale is not the same as evidence covering it
The 2003 edition requires Type I scales to be marked with a zero position, but the presence of zero does not make the specified range start from zero. Readers still need to go back to the maximum value, the applicable range, and the measurement records to confirm what the documents actually support.
If the torque you need is below 20 % of the maximum, a document that covers only the general 2003 range cannot by itself guarantee that the tool conforms at that working point. You can ask the supplier whether additional specifications and measurement evidence exist, then evaluate the intended use; or you can choose a tool whose specified range includes the working torque. This reading of the evidence does not claim that a tool is necessarily damaged or completely unable to operate below 20 %.