Under the 2003 edition, must a tool be recalibrated after repair or improper handling?
Applies to: ISO 6789:2003 · Updated 2026-11-26
Technical review: Alston Hsiao (Chieh-Jen Hsiao), CTO, STAND
Under the 2003 edition, a tool must be recalibrated after repair, after an overload greater than the values given in subclause 5.1.6, or after any improper handling that might influence the errors of measurement. These conditions are independent of the routine interval; even if the scheduled due date has not yet arrived, the tool’s measurement performance after the event must be dealt with.
A completed repair does not confirm the tool’s measurement performance
A repair record can show what was done, but it cannot by itself answer how the tool measures now. Subclause 5.3.2 of the 2003 edition lists “after repair” as a condition for recalibration and does not require proof that the tool has gone outside its tolerance before recalibration is started.
For improper handling, the key question is whether it might influence the errors of measurement. For example, if a tool has been struck and there is reason to suspect that its indicating or release mechanism has been affected, it must be assessed under this condition; the absence of obvious external damage cannot by itself prove that the measurement performance is unchanged. Nor does this example mean that every contact or movement automatically counts as the same kind of event.
Match the overload to that edition’s basis
The overloading test value in subclause 5.1.6 of the 2003 edition is 125 % of maximum capacity; fixed-adjustment Type II Classes B/E use nominal capacity as the basis. This overloading test does not apply to limiting torque tools, so the figure cannot be applied to every tool regardless of kind.
The following is teaching data: suppose a tool to which the maximum-capacity basis above applies has a maximum capacity of 80 N·m, and it is known to have been subjected to 108 N·m during an event.
Overload threshold = 80 × 125 ÷ 100 = 100 N·m.
Because 108 N·m > 100 N·m, this scenario meets the recalibration trigger of “greater than the values given in the clause.” The calculation is for identifying the event; it does not instruct users to apply test loads themselves, nor does it permit routine use above the applicable maximum capacity or nominal capacity.
Even without exceeding the figure, look at the event itself
The overload value is not the only condition. Even if the known load did not exceed the threshold above, there may still have been a repair, or improper handling that might influence the errors of measurement; if the peak value is unknown, it is not advisable to record simply “no overload, no action needed.” Record the known facts and have the people responsible for tool management and calibration confirm the situation.
When arranging recalibration, you can also provide the repair performed, what happened during the abnormal event, and the tool’s identification. The calibration still follows the 2003 method and tolerance requirements; in practice, linking the event record to the new results is what makes it possible to show which confirmation the later use is based on.