Which checks must be repeated after a tool is repaired or improperly handled?

Applies to: ISO 6789-1:2017, ISO 6789-2:2017 · Updated 2026-11-26

Technical review: Alston Hsiao (Chieh-Jen Hsiao), CTO, STAND

Key answer

After repair, an overload greater than the values given in Part 1, subclause 5.1.6, or any improper handling that might influence the tool’s performance and the fulfilment of the quality conformance requirements, the conformance test must be repeated in accordance with Part 1; if the tool is managed by calibration under Part 2, a calibration must also be arranged in accordance with the Part 2 event-triggered requirement. These requirements are triggered by events, so there is no need to wait until the routine interval expires; the actual work should clearly state which part it is based on.

First determine which kind of event occurred

“After repair” is itself a condition listed in the clause; it is not something to arrange only after trial use shows that the deviation has grown. For improper handling, the question is whether it might influence performance and the fulfilment of the quality conformance requirements—for example, whether an impact could have affected the mechanism. An intact appearance alone is not enough to rule this out.

Judging an overload requires going back to Part 1, subclause 5.1.6, and the conditions that apply to the tool. The classes that this subclause loads at nominal capacity are printed as Type II (Classes D, E and F); by classification, the fixed-adjustment tools should be Type II Classes B/E. This article follows that reading and retains the printed wording; this is not an official corrigendum. See Which capacity applies to fixed-adjustment tools in the ISO 6789-1 overloading test?. Where such tools are involved, first check the classification and capacity data to avoid applying the wrong basis.

The requirements of the overloading test cannot be treated as permission to overload a tool in everyday use either. Even if the peak value at the time cannot be reconstructed, you still need to assess whether improper handling that might influence performance was involved; it is not advisable to conclude that there was no effect simply because “no load above the threshold was measured.”

Keep conformance and calibration clearly separate

Work you need to completeBasis and purpose of the result
Confirm whether the tool still conforms to the requirementsPart 1, subclause 5.3, refers to subclause 5.2; measure as specified and make the conformance decision.
Obtain calibration results after the eventPart 2, subclause 4.1, triggers calibration after the event; the calibration method follows subclause 4.2, the measurement error is calculated under Clause 5, and the uncertainty is evaluated under Clauses 6 and 7.

The two kinds of work can be specified together when they are arranged, but do not treat “repair completed” or “appearance normal” as a test or calibration result. Nor does this mean repeating every design-stage test unchanged: the conformance test during use has its own specified method, and calibration has its own result and document requirements.

Keep later records connected

In practice, you can first label the tool as pending confirmation and record the repair performed, what happened during the abnormal event, any known loads, and the accessory configuration, then provide this to the people performing the test or calibration. This is a management suggestion intended to link the event and the new results to the same tool.

Before returning the tool to work, confirm that the tests, calibration, and decisions required for its intended use have been completed, then update the tool records. The standard’s event-triggered conditions do not by themselves tell you how far the event affected work done before it; if that needs to be traced, actual event and measurement data are still required.