Under ISO 6789:2003, how often should the tool be recalibrated?

Applies to: ISO 6789:2003 · Updated 2026-11-26

Technical review: Alston Hsiao (Chieh-Jen Hsiao), CTO, STAND

Key answer

Under ISO 6789:2003, when no procedure for the control of test devices is used, a period of use of 12 months or approximately 5000 cycles can be taken as the default recalibration interval. Where a control procedure exists, the interval must be adapted according to the conditions of use and the experience gained during recalibration. This is the arrangement in the 2003 edition, and the wording of the 2017 edition cannot be carried over into it wholesale.

First distinguish the default value from your own management interval

The 2003 edition treats torque tools as test devices to be managed. When choosing the interval, consider the accuracy required by the work, the frequency of use, the typical load, and the operating environment and storage conditions; then review the interval using the experience accumulated through recalibration.

So “send it out once a year” can be a management decision, but the standard alone does not make it the only answer that can be set for every tool. A frequently used tool and an occasionally used tool, even of the same model, each need their own usage and history of results to be tracked.

Without a control procedure, the original text gives “12 months or approximately 5000 cycles.” The cycle count is qualified by “approximately,” and the text does not contain the “whichever occurs first” that the 2017 edition states explicitly. If a user organization chooses the more conservative whichever-first approach, it should record this clearly as its own procedure rather than presenting the added wording as a sentence from the 2003 edition.

The first interval starts with the user’s first operation

For the first recalibration, the 2003 edition places the starting point at the user’s first operation of the tool. It is therefore best to keep separate records of a new tool’s purchase date, the date it entered storage, and the date it was put into service; the date on an earlier certificate alone is not enough to establish when the tool was first used.

Subsequent intervals must be reviewed and adapted in light of the control procedure and the experience gained during recalibration. There is no rule giving an extension factor that applies automatically whenever a calibration is passed, and starting the interval at first operation does not mean that a tool’s condition can be disregarded however long it is stored; storage conditions are themselves a factor in choosing the interval.

The interval should be traceable to records

In practice, the tool’s identification, the date it was put into service, its usage, typical loads, and successive calibration results can be kept in a single register, so that the basis of each scheduling decision can be explained. This is a convenient way of organizing the work, not a fixed form required by the standard.

The 2003 edition also requires applicable factory or legal regulations to be observed, so the default interval cannot be used to set other requirements aside. Recalibration is also required after repair, after an overload greater than the values given in subclause 5.1.6, or after any improper handling that might influence the errors of measurement. These are separate triggering conditions; such an event should not be left for the next routine schedule just because the calendar due date has not yet arrived.