How often should the tool be recalibrated?
Applies to: ISO 6789-2:2017 (with reference to ISO 6789-1:2017) · Updated 2026-11-26
Technical review: Alston Hsiao (Chieh-Jen Hsiao), CTO, STAND
Under ISO 6789-2:2017, if no procedure for the control of test devices is used, a recalibration interval of 12 months or 5000 cycles, whichever occurs first, may be adopted. Where a control procedure exists, torque tools must be included in it, and the interval must be set and adjusted according to the conditions of use and the results of successive calibrations; not every tool is simply recalibrated once a year.
The default interval starts with first use
The Part 2 default interval starts with the first use of the tool. The date of receipt, the certificate date, and the date of first use should therefore be recorded separately. If a tool is placed in storage before being put into service, the management records need to show when it was actually put into use, rather than treating the date on the certificate as the starting point in every case.
When these default values are adopted, both time and number of cycles need to be watched. A frequently used tool may reach the cycle threshold first; for an infrequently used tool, a time threshold that has been reached should not be overlooked merely because its cycle count is low. This is what “whichever occurs first” means for scheduling.
Without a record of cycles, the calendar alone cannot show that the cycle limit has not yet been reached. In practice, a tool register can record when the tool was put into service, its usage, and its calibration records, so that the schedule has a basis that can be checked. This is a suggested management practice, not a form layout specified by the standard.
With a control procedure, review the interval using results
When setting the interval, consider the maximum permissible measurement error required for the work, the frequency of use, the typical load, and the operating and storage conditions. Then use the results of several successive calibrations to review whether the original interval is still appropriate.
For example, if the tool’s frequency of use changes, the original schedule is worth re-evaluating; but a single good result does not justify extending the interval indefinitely. The standard provides a principle of adjusting the interval through the control procedure and the results; there is no automatic extension factor that applies to every tool.
If the user, their customer, or legislation requires a shorter interval, the shorter interval may be used and built into the schedule. This article does not assume that any particular customer requires a specific number of months; the actual requirements still need to be confirmed.
Schedule recalibration and conformance testing separately
Part 1 also specifies conformance testing during use, whose purpose is to confirm whether the tool still conforms to the requirements; recalibration under Part 2 provides calibration results and their uncertainty. A management register can track both together, but it should state clearly which kind of work is being scheduled.
In addition, repair, an overload greater than the values given in Part 1, subclause 5.1.6, or any improper handling that might influence the tool’s performance and the fulfilment of the quality conformance requirements separately triggers the Part 1 conformance test and the Part 2 calibration. In these situations, you cannot keep following the original schedule merely because the routine due date has not yet arrived.