How do you confirm that a tool still conforms to Part 1 after a period of use?

Applies to: ISO 6789-1:2017 · Updated 2026-11-26

Technical review: Alston Hsiao (Chieh-Jen Hsiao), CTO, STAND

Key answer

To confirm that a tool in use still conforms, carry out the Part 1 conformance test during use and check that each result is still within the applicable maximum permissible relative deviation. The basis is subclause 5.3, which refers to subclause 5.2, with the measurements then completed in accordance with Clause 6; an original conformance record from manufacture cannot be taken to show that the tool has remained in conformance ever since.

First state clearly what this test is to confirm

This test is concerned with whether the tool still conforms to the requirements after a period of use. Before arranging it, match the tool’s identity, Type and Class, specified range, measurement directions, and the applicable permissible deviation. If the manufacturer claims a maximum permissible relative deviation smaller than the standard’s, the results must also be judged against that claimed value.

For example, an old declaration issued at manufacture shows which measurements were made at that time; to answer what the tool’s condition is now, you still need the results of this test. Matching the old and new records to the same tool under corresponding conditions also helps when reviewing the test interval later.

Base the decision on complete measurement results

The Part 1 measurement arrangements cover the applicable measurement system, the environmental and mounting conditions, preloading, and the measurement point and reading sequence for the tool’s class. A quick pull at a single torque point on the shop floor may help reveal a problem, but it cannot be used to claim that the conformance test required by the standard has been completed.

After measurement, calculate each relative deviation using the Part 1 formula and compare it with the applicable maximum permissible relative deviation. The requirement applies to every result; it is not enough for the mean of the readings to fall within the limit. When this conformance decision is made under subclause 5.1.5, the influence of the uncertainty of the tool and of the torque measurement device is not considered in the decision.

If the work also requires traceable calibration results and uncertainty for this tool, separately confirm whether the Part 2 calibration requirements have been completed. This is a different evidential need; seeing the word “conforms” alone is not enough to confirm that the required calibration information has been provided.

Include the test in the management of tools in use

Organizations that already have procedures for the control of test devices must include torque tools in them and adjust the arrangements according to the conditions of use and the results of successive conformance tests. Records should retain the tool’s identification, directions, range, and measurement results so that later reviews can find the data for the same tool.

Part 1 also allows the user to define a different specified torque range for the tool; if a declaration of conformance is issued, that range must be stated in accordance with Clause 8 e), and this cannot be taken to mean that the manufacturer’s original performance assurance has been extended. How to schedule the interval, and what to do after repair or an abnormal event, are explained in the dedicated articles below.