How should a catalog’s “accuracy ±4 %” be understood under the standard?
Applies to: ISO 6789-1:2017 · Updated 2026-11-17
Technical review: Alston Hsiao (Chieh-Jen Hsiao), CTO, STAND
For “accuracy ±4 %” in a catalog, first confirm whether it is claiming a maximum permissible relative deviation, and under what conditions the claim applies. In ISO 6789-1:2017, accuracy is not a quantity that is given a numerical value directly; what can be compared result by result is the relative deviation, which has a defined basis and defined limits.
Map the sales term back to a checkable specification
“Accuracy” is a concept describing how good a measurement is: the smaller the deviation, the more accurate the measurement is said to be. It is not itself a quantity to which a number can be assigned, yet everyday usage often pairs it with a percentage. Subclause 3.1.4 of Part 1 specifically distinguishes this customary usage from the maximum permissible deviation, so that readers do not treat quantities with similar names as the same thing.
In the way Part 1 evaluates results, what matters is how far the observed value from the measurement device deviates from the tool’s target value. The maximum permissible relative deviation sets the acceptable limits for that percentage. If a catalog’s ±4 % is meant to correspond to this requirement, it should state the part, year, and conditions on which it is based.
From a bare “accuracy ±4 %,” therefore, you cannot infer what the percentage is based on, and still less can it be taken directly as the uncertainty on a calibration certificate. First confirm which quantity it describes; only then can the number be meaningfully compared.
Check the scope of the claim
When confirming with the supplier, you can focus your questions on three points:
- Basis and classification: Is it based on ISO 6789-1:2017? Which Type and Class is the tool, and what is its maximum torque?
- Conditions of application: Which torque range and operating directions does the claim cover? Are the tool and accessory configuration you have within the scope described?
- Basis of evaluation: Does ±4 % mean the maximum permissible relative deviation calculated in accordance with Part 1? What limit does the declaration of conformance actually use?
These are questions for checking specifications; they do not mean the standard requires catalogs to use any fixed wording. If the supplier uses a different definition, keep that definition and confirm whether it meets your needs, rather than rewriting it yourself as an ISO conformance claim.
A limit is not this tool’s measured result
The maximum permissible relative deviation is a decision limit, whereas the actual relative deviation comes from each individual measurement. Tools marked with the same ±4 % still need their own observed values before you can know how far they deviated when measured and whether they were within the limit.
Part 1 requires every result to meet the applicable limit; if the manufacturer claims a value smaller than the tabulated value, every result must also meet that claim. When receiving tools, checking the catalog and the declaration of conformance together is more useful than looking only at the word “accuracy.” A single catalog phrase is not enough either to confirm that a product conforms or to decide that it does not.