If the mean result is within tolerance, does every measurement conform?
Applies to: ISO 6789-1:2017 · Updated 2026-11-23
Technical review: Alston Hsiao (Chieh-Jen Hsiao), CTO, STAND
A mean result within tolerance does not mean that every measurement meets the requirement; Part 1 requires each result to be judged individually. Higher and lower results can cancel each other out when averaged. To confirm conformance with the maximum permissible relative deviation, you still have to go back to each observed value.
What the standard requires to be checked
Subclause 5.1.5 of ISO 6789-1:2017 requires every result recorded and calculated according to the measurement procedure to fall within the limit applicable to the tool. Where the manufacturer claims a smaller value, every result must also meet that claimed value.
A mean can therefore be calculated in addition when organizing the data, but it cannot be the only basis for the decision. The mean shows where the center of the data set lies, while judging each result individually preserves the fact that one measurement departed too far. The two serve different purposes. If a report presents only a single mean, readers cannot check whether any individual result exceeded the limit.
The mean is within the limit, but the last result is not
The following is teaching data: suppose a Type II Class A wrench has a specified range of 20–100 N·m, the applicable tolerance is ±4 %, and no smaller value is claimed. In one direction, at a target value of 60 N·m, the 5 records specified for that point are completed.
Each result is calculated with a_d = (X_o − 60) ÷ 60 × 100:
| No. | Observed value X_o (N·m) | Relative deviation calculation | Within ±4 %? |
|---|---|---|---|
| 1 | 58.2 | (58.2 − 60) ÷ 60 × 100 = −3.0 % | Yes |
| 2 | 59.4 | (59.4 − 60) ÷ 60 × 100 = −1.0 % | Yes |
| 3 | 60.0 | (60.0 − 60) ÷ 60 × 100 = 0.0 % | Yes |
| 4 | 60.6 | (60.6 − 60) ÷ 60 × 100 = +1.0 % | Yes |
| 5 | 63.0 | (63.0 − 60) ÷ 60 × 100 = +5.0 % | No |
The mean of the observed values is (58.2 + 59.4 + 60.0 + 60.6 + 63.0) ÷ 5 = 60.24 N·m.
Calculating with the mean alone gives (60.24 − 60) ÷ 60 × 100 = +0.4 %, which appears to be within the limit. But the last result is +5.0 %, so it cannot be claimed that every result in this set conforms. This illustrates a Part 1 comparison that uses the mean at the same target value; it is not the Part 2 method for the mean relative measurement error.
Keep the ability to check individual results when receiving tools
When you see a mean, look next for the original observed values, the corresponding target values, and the tolerance used. A declaration of conformance must also give the observed values for each target torque and a statement of whether they lie within the permissible deviation.
This example shows one result outside the limit, but it does not supply the cause. Review the measurement procedure, the condition of the tool, and the data handling against the actual records before deciding on follow-up action. This teaching data alone cannot diagnose a damaged tool, and the result outside the limit must not be deleted so that only the mean of the remaining data is reported.