How is the relative deviation calculated in Part 1?

Applies to: ISO 6789-1:2017 · Updated 2026-11-23

Technical review: Alston Hsiao (Chieh-Jen Hsiao), CTO, STAND

Key answer

In Part 1, the relative deviation is the observed value from the measurement device minus the tool’s target value, divided by the target value and multiplied by 100 %. The formula is a_d = (X_o − X_t) ÷ X_t × 100. The denominator is the target value, and each observed value is calculated separately.

First, identify the two torque values

X_t is the target value on the tool side; X_o is the observed value shown by the measurement device. Both are entered in the same torque unit, and the resulting a_d is expressed as a percentage.

How the target value is obtained depends on the kind of tool and must be confirmed. For indicating tools, use the tool’s indicated value; for Type II Class A, D, and G setting tools, use the value set on the scale or display. Under subclause 7.1, fixed-adjustment Classes B and E use the nominal set value, and non-graduated adjustable Classes C and F use the lowest limit value or the nominal pre-set value. You cannot assume that every tool has a setting scale that can be read.

The easiest mistake is to put the device reading in the denominator. Part 1 asks how far the observed result departs from the original target, so the target value is the basis of the ratio.

Calculating a single result

The following is teaching data: a Type II Class A wrench has a specified range of 20–100 N·m, and the manufacturer makes no claim of a smaller deviation. It is set to 60 N·m, and at release the measurement device observes 61.8 N·m.

First assign the role of each quantity: X_t = 60 N·m and X_o = 61.8 N·m. The calculation is:

a_d = (61.8 − 60) ÷ 60 × 100

= 1.8 ÷ 60 × 100 = +3.0 %

The plus sign means that the torque observed by the measurement device is higher than the tool’s target value. If the observed value is lower than the target value, the same order of subtraction gives a negative value. The sign expresses the relative size of the two torque values; it is not a code for clockwise or counterclockwise operation.

Interpreting the percentage

In this example the maximum torque is 100 N·m, so the tabulated tolerance for Type II Class A is ±4 %. Because +3.0 % lies within that limit, this result conforms to the tolerance. If the manufacturer claims a smaller value, check against the applicable claim instead.

One conforming result does not mean that every measurement point and direction of the tool conforms. Keep the original observed values and calculate each result obtained by the specified procedure individually. This also makes later checks easier: was the tool’s target value entered incorrectly, was the wrong denominator used, or did a result genuinely exceed the limit?

If a report is based on Part 2, use that part’s formula for the relative measurement error; giving this percentage a different column name is not enough.