Are the percentage formulas in Part 1 and Part 2 interchangeable?
Applies to: ISO 6789:2003, ISO 6789-1:2017, ISO 6789-2:2017 · Updated 2026-11-23
Technical review: Alston Hsiao (Chieh-Jen Hsiao), CTO, STAND
They are not directly interchangeable: the percentages in Part 1 and Part 2 differ in both the order of subtraction and the denominator. The same pair of tool and device values can give one positive and one negative result, with different absolute values. Before switching formulas, confirm the document basis and the role of each number.
The basis of each formula
The Part 1 relative deviation, a_d = (X_o − X_t) ÷ X_t × 100, compares the observed value from the measurement device with the tool’s target value; the denominator is the target value.
The Part 2 relative measurement error, a_s = (X_a − X_r) ÷ X_r × 100, compares the target value on the tool side with the device’s reference value; the denominator is the reference value.
So even when, in a particular measurement situation, X_t and X_a take the same number and X_o and X_r take the same number, the basis of the ratio is still different. Changing the column names, or multiplying the result by −1, does not complete a conversion.
How the results differ for the same teaching data
The following is teaching data: suppose an indicating tool shows 60 N·m, and the device value in the same measurement is 61.8 N·m. Only this one pair of values is used, to compare the formulas; no claim is made that the full procedure of any edition has been completed.
| Document and quantity | Values entered | Result |
|---|---|---|
| 2017 Part 1 relative deviation a_d | (61.8 − 60) ÷ 60 × 100 | 1.8 ÷ 60 × 100 = +3.0 % |
| 2017 Part 2 relative measurement error a_s | (60 − 61.8) ÷ 61.8 × 100 | −1.8 ÷ 61.8 × 100 ≈ −2.913 % |
| 2003 edition deviation A_s | (60 − 61.8) ÷ 61.8 × 100 | −1.8 ÷ 61.8 × 100 ≈ −2.913 % |
For the same measurement, the plus sign in Part 1 means that the device’s observed value is higher than the tool’s target value; the minus sign in Part 2 means that the tool-side value is lower than the reference value. Both statements describe the same relationship in size, so they do not contradict each other. The absolute values differ because one is divided by 60 and the other by 61.8.
Handling 2003 data and spreadsheets
The deviation formula in ISO 6789:2003 subtracts the calibration device’s reference value from the tool value and divides by the reference value, so in this example it gives the same number as Part 2. But this is only a comparison of formulas in the chosen situation; it does not follow that the two editions have the same overall calibration procedure, the same definitions of target value, or the same document requirements.
When checking a spreadsheet, keep the original tool values, device values, edition, and quantity names, then recalculate according to the part required. In particular, for the target values of non-graduated tools, go back to the requirements of each edition; the direct correspondence in this example cannot be extended to every Class.
When making a decision, the formula should also match the purpose: the Part 1 maximum permissible relative deviation should be compared with an a_d calculated according to Part 1. An error percentage copied from a Part 2 certificate cannot be used as a direct substitute, even if the numbers are close.