What do w, W, and W′ mean?
Applies to: ISO 6789-2:2017 · Updated 2026-11-23
Technical review: Alston Hsiao (Chieh-Jen Hsiao), CTO, STAND
w is the relative standard measurement uncertainty; W is the quantity obtained by expanding it with the coverage factor; and W′ further adds the absolute value of the tool’s mean relative measurement error and the absolute value of the stated relative measurement error of the measurement device, that is, W′ = |ā_s| + W + |b_ep|. All three are expressed as percentages, but they answer different questions. When reading a certificate, keep the upper and lower case and the prime mark as written.
Find the calculation by the role of the quantity
| What you are checking | Role of the quantity and where to read more |
|---|---|
| How are the individual effects combined? | w is the tool’s relative standard measurement uncertainty at that calibration point. See How are the uncertainty components combined into the tool’s w?. |
| Why multiply by a coverage factor as well? | W is obtained by expanding w with the coverage factor. See Is multiplying w by 2 enough to get W?. |
| Why does the interval include more than W? | W′ also includes the absolute value of ā_s and the absolute value of b_ep. See Why does W′ also include the mean error and the device error?. |
| Which mean is the mean error? | ā_s is the mean of the individual relative measurement errors. See Can the mean relative measurement error be calculated directly from the mean reference value?. |
| How is the device error turned into a percentage? | b_ep is the stated relative measurement error of the measurement device, used in calculating the interval. See How is the device error b_e converted into the b_ep used in tool calculations?. |
| At what point is rounding applied? | Rounding is required before the components are combined and before w is expanded. See At which step must uncertainty calculations be rounded to three decimal places?. |
Keep the calibration point and the object in view on a certificate
A Part 2 certificate is required to list W and W′ for each calibration point of the tool; w is an important intermediate quantity for understanding the calculation. If a table shows only a single “uncertainty” column, first find out which quantity it refers to before comparing it with other documents.
The same set of data may also contain W_md or W′md for the measurement device. The subscript indicates a different object: the device’s result is not the result for the tool being calibrated. A tool calibration must include the applicable components in its evaluation; a percentage cannot simply be copied from the device’s certificate.
Nor are these three quantities three grades of conformance. The terms that W′ includes beyond W each have their own values and sources; before interpreting them, match them to the tool, direction, and calibration point. Each calculation and the rounding are covered in the articles linked in the table above, so that conformance is not guessed from whether a symbol is upper- or lower-case, or from the size of the value alone.