When judging conformance under Part 1, should uncertainty be added to the deviation?
Applies to: ISO 6789-1:2017 · Updated 2026-11-23
Technical review: Alston Hsiao (Chieh-Jen Hsiao), CTO, STAND
When judging conformance with the maximum permissible relative deviation under subclause 5.1.5 of Part 1, the uncertainty of the tool and of the measurement device is not taken into account: it is not added to the deviation, and it is not used to narrow or widen the limit. First calculate each relative deviation with the Part 1 formula, then compare it directly with the applicable limit. Any decision rule separately agreed with a customer needs to have its basis and purpose stated separately.
Which quantity this decision compares
The relative deviation used in ISO 6789-1:2017 is a_d = (X_o − X_t) ÷ X_t × 100. X_t is the tool’s target value and X_o is the value observed by the measurement device; each calculated percentage is compared with the tool’s maximum permissible relative deviation.
Subclause 5.1.5 states explicitly that the effect of the uncertainty of the tool and the measurement device is not considered when judging conformance with that subclause. This does not mean that the uncertainty is zero, nor that any measurement device may be used. The device must still meet the capability and traceability requirements of Part 1, and the declaration of conformance must still give information on the device’s error and uncertainty interval.
The same method applies close to the limit
The following is teaching data: a Type II Class A wrench has a specified range of 20–100 N·m, the manufacturer makes no claim of a smaller deviation, and the applicable tolerance is ±4 %. The target value is 60 N·m, and a single observed value is 62.28 N·m.
a_d = (62.28 − 60) ÷ 60 × 100
= 2.28 ÷ 60 × 100 = +3.8 %
This result lies within −4 % to +4 %, so it conforms as far as the deviation decision under subclause 5.1.5 is concerned. Even though the result is close to the upper limit, do not add a separately stated uncertainty to it on your own initiative and then call the new sum the Part 1 relative deviation. Likewise, do not use the uncertainty to narrow or widen the ±4 % limit.
This is the interpretation of a single result only; the other measurement points, directions, and individual records still need to be confirmed. If the manufacturer claims a smaller limit, compare each result with that claimed value instead.
Keep Part 2’s W′ and separately agreed rules apart
In Part 2, W′ is the relative measurement uncertainty interval. It is made up of the absolute value of the mean relative measurement error, the relative expanded measurement uncertainty W, and the absolute value of the stated relative measurement error of the measurement device (b_ep). It is not the single-result a_d of Part 1; even though both are expressed as percentages, one cannot simply be moved into the other’s column.
If a customer separately agrees that uncertainty is to be included in the acceptance decision, clearly record which quantities, which limits, and which way of reaching the decision are used, and report that acceptance conclusion separately from the conclusion under subclause 5.1.5 of Part 1. This is a way of avoiding confusion between the two kinds of requirement. It does not mean that every customer must set a separate rule, and this article does not make a decision on any contract.